Commands
TDS5000B Series Service Manual
5-31
<NR1> selects an area by number, which can range from 0 (zero selects all)
through 15 (as specified by DIAg:NUMITEMS).
DIAg:SELect:AREA 1
This command selects Area 1 as the area to be diagnosed.
DIAg:SELect:LAST
This command (no query form) selects one or more diagnostic menu items to be
executed via the DIAg:EXECUTE command. If you specify DIAg:LEVEL:SUB-
SYS, then menu items come from this diagnostic level and are limited to the
value returned by the DIAg:NUMITEMS? query. For example, specifying
DIAg:SELECT:SUBSYS 2 and DIAg:NUMITEMS 5 indicates that diagnostics
will st art from subsystem 2 and that you can specify a range from 2 through 5 for
DIAg:SELect:LAST.
If you enter:
DIAg:SELect:LAST 2, only subsystem 2 will be executed.
DIAG:SELect:LAST 4, subsystems 2 through 4 will be executed.
DIAg:EXECUTE, DIAg:LEVEL, DIAg:NAMe:AREA?, DIAg:NAMe:SUB-
SYS?, DIAg:NAMe:TEST?, DIAg:NUMITEMS?, DIAg:SELect:SUBSYS
DIAg:SELect:LAST <NR1>
<NR1> selects an integer that identifies the number of the last item that will be
executed when the DIAg:EXECUTE command is run.
DIAg:SELect:LAST 2
specifies that (based on the previous example, only subsystem 2 wil l be
executed.
DIAg:SELect:SUBSYS
This command (no query form) selects one of the available subsystems. This
command is equivalent to selecting Instrument Diagnostics from the Utilities
menu and then choosing a Subsystem from the drop-down list.
Arguments
Example
Related Commands
Syntax
Arguments