Commands
TDS5000B Series Service Manual
5-33
DIAg:STATE
This command (no query form) changes the oscilloscope operating state.
Depending on the argument, diagnostics capabilities are either turned on or off.
This command i s equivalent to opening the DIAg:STATE dialog (ON) or closing
it (OFF).
TEST:STOP
DIAg:STATE {ON|OFF}
ON puts the instrument into the state in which diagnostics can be run. This
argument is used automatically if either DIAg:EXECUTE or DIAg:TEST
commands are executed.
OFF disables diagnostics capabilities and returns the oscilloscope to a normal
operating state.
DIAg:STATE OFF
turns off diagnostics capabilities and returns the instrument to a normal operating
state.
DIAg:STOP
This command (no query form) causes diagnostics (or test) execution to
terminate at the end of the next l ow-level test. This command is equivalent to
selecting Instrument Diagnostics from the Utilities menu and then clicking
Abort.
TEST:STOP
DIAg:STOP
DIAg:STOP
terminates diagnostics (or test) execution at the end of the next low-level test.
Related Commands
Syntax
Arguments
Example
Related Commands
Syntax
Example