Commands
5-34
TDS5000B Series Service Manual
TEST
This command (no query form) provides the ability t o select and execute an item
at any level of the test hierarchy (Test, Area or Subsystem). This command is
equivalent to selecting Instrument Diagnostics from the Utilities menu, choosing
a test and then pressing Run.
NOTE. Turning off both DIAg:CONT R OL:HALT and DIAg:CONTROL:LOOP
before executing the TE ST command is recommended.
TEST <NR3>
<NR3> sets the test ID, which ranges from 0 through 3 characters. If no test ID
is specified, all available diagnostics are executed.
TEST ”1”
executes all Acquisition tests (Subsyst em: Acquisition)
TEST ”11”
executes all Acquisition Memory tests (Subsyst em: Acquisition, Area:Memory)
TEST ”113”
executes a specific Acquisition Memory test (Subsystem: Acquisition, Area:
Memory, Test: Address Lines).
TEST:RESults?
This query-only command returns an abbreviated status about the results of the
last TEST execution. Thi s command is equivalent to selecting Instrument
Diagnostics from the Utilities menu and then reviewing the Diagnostic Status.
TEST, TEST:RESults:VERBose?
TEST:RESults?
TEST:RESult s?
Syntax
Arguments
Example 1
Example 2
Example 3
Related Commands
Syntax
Example 1