Commands
TDS5000B Series Service Manual
5-35
might return
:TEST:RESULTS ”135”
indicating the specific failure code of the first test failure.
TEST:RESult s?
might return
:TEST:RESULTS ”PASS”
indicating the previously executed t est passed
TEST:RESult s?
might return
:TEST:RESULTS ”*****”
indicating the test was not run.
TEST:RESults:VERBose?
This query-only command returns a more explanatory message about the results
of the last TEST execution than the TEST:RESults query.
TEST, TEST:RESults?
TEST:RESults:VERBose?
TEST:RESul t s: VE RBose?
might return
:TEST:RESULTS:VERBOSE DIAG ABORTED, ERROR 15 in A2D-Dmux 1