Sample Calibration Program
5-42
TDS5000B Series Service Manual
End If
End Sub
Function to wait for Operation Complete to happen
from the tekScope.
Private Function tekScopeOPC() As Boolean
Dim opc As Integer
Dim strOPC As String
Dim count As Integer
On Error Resume Next
opc = 0
count = 0
While (opc <> 1) And (count < 1000)
Sleep 1000
strOPC = ””
strOPC = TvcScope.Query(”*OPC?”)
If Len(strOPC) > 0 Then
opc = Asc(strOPC) - Asc(”0”)
Else
opc = 0
End If
count = count + 1
Wend
If count > 10 Then
Debug.Print ”OPC happened after & count & seconds.”
End If
If opc <> 1 Then
Debug.Print ”Timed out trying to get OPC on tekScope.”
tekScopeOPC = False
Else
tekScopeOPC = True
End If
End Function
Resets the WaveTek 9500 to default mode.
Private Sub resetWaveTek()
Dim count As Integer
Dim opc As Integer
Dim strOPC As String
Dim esr As Integer
TvcWaveTek.WriteString ”*CLS”
TvcWaveTek.WriteString ”*RST”
count = 0
strOPC = TvcWaveTek.Query(”*OPC?”)
If Len(strOPC) > 0 Then
opc = Asc(strOPC) - Asc(”0”)
Else
opc = 0