Sample Calibration Program
5-44
TDS5000B Series Service Manual
End If
TvcWaveTek.WriteString strCommand
strCommand = ”ROUTE:SIGNAL:PATH CH” & reqChannel
TvcWaveTek.WriteString strCommand
strCommand = ”ROUTE:SIGNAL:IMPEDANCE & reqTermination
TvcWaveTek.WriteString strCommand
Case 4
Debug.Print ”use deskew signal”
TvcWaveTek.WriteString ”SOUR:FUNC:SHAPE SKEW”
TvcWaveTek.WriteString ”SOUR:PAR:SKEW:ALIG PREC”
TvcWaveTek.WriteString ”SOUR:FREQ:FIX 1000000”
TvcWaveTek.WriteString ”ROUT:SIGN:SKEW CH1, ON”
TvcWaveTek.WriteString ”ROUT:SIGN:SKEW CH2, ON”
TvcWaveTek.WriteString ”ROUT:SIGN:SKEW CH3, ON”
TvcWaveTek.WriteString ”ROUT:SIGN:SKEW CH4, ON”
TvcWaveTek.WriteString ”ROUT:SIGN:SKEW CH5, ON”
Case Else
Debug.Print ”use AC signal”
TvcWaveTek.WriteString ”SOUR:FUNC:SHAPE SIN”
strCommand = ”SOURCE:FREQ & reqFreq
TvcWaveTek.WriteString strCommand
strCommand = ”SOUR:VOLT & reqVolts
TvcWaveTek.WriteString strCommand
strCommand = ”ROUTE:SIGNAL:PATH CH” & reqChannel
TvcWaveTek.WriteString strCommand
strCommand = ”ROUTE:SIGNAL:IMPEDANCE & reqTermination
TvcWaveTek.WriteString strCommand
End Select
TvcWaveTek.WriteString ”OUTPUT ON”
Sleep 1000
End Sub
Subroutine parses the information from the TDS5000B as to the required signal
for the next calibration step.
Private Sub updateSignal(strRequired As String)
Dim strSignal() As String
Dim strMessage As String
Dim stepStrLen As Integer
Dim stepStr As String text progress bar
If strRequired = strEndOfCal Then
lstOutput.AddItem ”Calibration complete”
calInitDone = False
Exit Sub
End If
strSignal = Split(strRequired, ”,”)
reqVolts = Mid(strSignal(0), 2)
reqFreq = strSignal(1)