Sample Calibration Program
TDS5000B Series Service Manual
5-45
reqChannel = strSignal(2)
reqTermination = strSignal(3)
reqStepName = strSignal(4)
strMessage = reqVolts & volts
If reqFreq = 0 Then
strMessage = strMessage & DC
Else
strMessage = strMessage & reqFreq & AC
End If
strMessage = strMessage & ”channel & reqChannel&””
If reqTermination = 1000000 Then
strMessage = strMessage & ”1 Meg Ohms”
Else
strMessage = strMessage & reqTermination & Ohms”
End If
stepStrLen = 2
stepStr = ”[]”
If Left(reqStepName, 3) = ”S01” Then
stepStrLen = 24
stepStr = ”[ long operation ]”
End If
If Left(reqStepName, 3) = ”D01” Then
stepStrLen = 10
stepStr = ”[ wait ]”
End If
txtSignal.Text = ”[” & String(stepNumber, ”+”) & stepStr & String(110 - (stepNumber + step-
StrLen), ”.”) & ”]”
txtSignal.Refresh
stepNumber = stepNumber + 1
If Left(reqStepName, 3) = ”S01” Then
stepNumber = stepNumber + 23
End If
If Left(reqStepName, 3) = ”D01” Then
stepNumber = stepNumber + 9
End If
lstOutput.AddItem ”Provide the following signal: & strMessage
lstOutput.ListIndex = lstOutput.ListCount - 1
lstOutput.Refresh
End Sub
Subroutine turns on the correct buttons for the current mode of operation at
the end of the calibration step.
Private Sub turnOnButtons()
cmdPlay.Enabled = True
cmdExit.Enabled = True
End Sub