Sample Calibration Program
5-46
TDS5000B Series Service Manual
Subroutine to turn off the buttons that will not work during the current
calibration step.
Private Sub turnOffButtons()
cmdPlay.Enabled = False
cmdExit.Enabled = False
End Sub
Subroutine to handle the Exit button.
This button is only available during single step mode or before calibration
has started.
Private Sub cmdExit_Click()
End
End Sub
Subroutine to handle the Next step.
Private Sub cmdNextStep()
Dim strResult As String
turnOffButtons
setWaveTek
TvcScope.WriteString ”:CALIBRATE:FACTORY CONTINUE”
If tekScopeOPC = False Then
Debug.Print ”factory calibration step did not complete”
Exit Sub
End If
Sleep 1000
strResult = TvcScope.Query(”:CALIBRATE:FACTORY:STEPSTATUS?”)
If Left(strResult, 4) = ”PASS” Then
lstOutput.AddItem ”Passed step + reqStepName
calPassedLastStep = True
Sleep 1000
strResult = TvcScope.Query(”CALIBRATE:FACTORY:STEPSTIMULUS?”)
updateSignal (strResult)
Debug.Print strResult
Else
If calPassedLastStep = False Then
calFailed = True
calInitDone = False
Else
Debug.Print ”Will retry last step”
End If
lstOutput.AddItem ”Failed step + reqStepName
calPassedLastStep = False
End If
lstOutput.ListIndex = lstOutput.ListCount - 1
turnOnButtons
End Sub
Subroutine to handle the Play button.
This will place the TDS5000B in factory calibration mode and if single step
mode is not active repeatedly do next step until calibration is complete.