Commands
5-10
TDS5000B Series Service Manual
CALibrate:CALProbe:CH2?
might return
:CALIBRATE:CALPROBE:CH2 0
indicating that the probe calibration passed for Channel 2.
CALibrate:FACtory
This command st arts t he factory calibration. This command functions the same
as the CALibrate:FACtory:STARt command.
CALibrate:FACtory
CALIBRATE:FACTORY
starts the factory calibration
CALibrate:FACtory:ABOrt
This command st ops the factory calibration, resets to t he initialization state, and
exits without changing the calibration data.
CALibrate:FACtory:STARt
CALibrate:FACtory:ABOrt
CALIBRATE:FACTORY:ABORT
stops the factory calibration
CALibrate:FACtory:CONTInue
This command does the next step in the calibration procedure. The calibration
step number is increased on successful completion of t hi s step. If the calibration
step fails, t he step number is not increased. You can use CALibrate:FACto-
ry:CONTinue, to repeat a factory calibration step that failed.
Follow this command with CALibrate:FACtory:STEPstatus? to determine
execution status.