Commands
TDS5000B Series Service Manual
5-11
If the current step is the last st ep and the calibration succeeds, new calibration
data and pass status is saved to non-volatile memory and the oscilloscope returns
to normal operation in the same state it was in before beginning factory
calibration.
If the current step is the last s t ep and any factory calibration st ep has failed,
without a successful pass for repeated steps, a fail status will be saved to
non-volatile memory. Factory calibration data will not be saved to non-volatile
memory if the factory status is FAIL.
CALibrate:FACtory:STARt
CALibrate:FACtory:CONTinue
CALIBRATE:FACTORY:CONTINUE
causes the instrument to execute the next step in the calibration procedure.
CALibrate:FACtory:FANcontrol
This command sets or returns the status of the fan control.
CALibrate:FACtory:FANcontrol <NR3> [, <NR3>]
<NR3> sets the temperature in degrees Celsius
<NR3> sets the mean sea level specified in feet; this command i s optional and if
not specified, it defaults to 0.0 (sea level)
CALibrate:FACtory:FANcontrol?
0 means that the fan was never calibrated
1 means that calibration is in process
2 means that the fan has been calibrated
CALibrate:FACtory:FANcontrol 20
This sets the fan control calibration temperature to 20 degrees Celsius at sea
level.
CALibrate:FACtory:FANcontrol?
Related Commands
Syntax
Example
Syntax 1
Arguments
Syntax 2
Returns
Example 1
Example 2