EasyManua.ls Logo

Teledyne 400E - External Digital I;O; Status Outputs; Figure 6-4: Status Output Connector

Teledyne 400E
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Instructions Model 400E Ozone Analyzer
Instruction Manual
112 P/N 04316 Rev: B
6.8. External Digital I/O
6.8.1. Status Outputs
The Status Outputs report analyzer conditions via optically isolated NPN transistors
which sink up to 50 mA of DC current. These outputs can be used interface with
devices that accept logic-level digital inputs, such as Programmable Logic
Controllers (PLC’s). Each Status bit is an open collector output.
NOTE
Most PLC’s have internal provisions for limiting the current that
the input will draw from an external device. When connecting to
a unit that does not have this feature, a 120 ohm external
dropping resistors must be used to limit the current through the
transistor output to 50mA or less. Refer to the Mainboard
Schematic 04070 in Appendix D.
The status outputs are accessed via a 12 pin connector on the analyzer’s rear panel
labeled STATUS. The function of each pin is defined in Table 6-11.
Connect to Internal
Ground of Monitoring
STATUS
1 2 3 4 5 6 7 8 D
+
SYSTEM OK
HIGH RANGE
CONC VALID
ZERO CAL
SPAN CAL
DIAGNOSTIC MODE
LOW SPAN
Figure 6-4: Status Output Connector

Table of Contents