Operating Instructions Model 400E Ozone Analyzer
Instruction Manual
148 P/N 04316 Rev: B
The Channel Properties, Triggering Events and Data Parameters/Functions for the
default channels are:
NAME:
CONC
EVENT: ATIMER
REPORT PERIOD: 000:01:00
NO. OF RECORDS: 800
RS-232 REPORT:
OFF
CHANNEL ENABLED: ON
NAME: O3REF
EVENT: ATIMER
REPORT PERIOD: 001:00:00
NO. OF RECORDS:
730
RS-232 REPORT: OFF
CHANNEL ENABLED: ON
OFF
NAME: PNUMTC
EVENT:
ATIMER
REPORT PERIOD: 001:00:00
NO. OF RECORDS: 360
RS-232 REPORT:
OFF
CHANNEL ENABLED: ON
NAME: O3GEN
EVENT: ATIMER
REPORT PERIOD:
001:00:00
NO. OF RECORDS: 360
RS-232 REPORT: OFF
CHANNEL ENABLED:
ON
NAME: CALDAT
EVENT: SLPCHG
REPORT PERIOD:
000:01:01
NO. OF RECORDS:200
RS-232 REPORT: OFF
CHANNEL ENABLED: ON
PARAMETER: CONC1
MODE: AVG
PRECISION:
1
STORE NUM SAMPLES
OFF
PARAMETER: PHREF
MODE: AVG
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: SMPLFLW
MODE: AVG
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: SMPLPRS
MODE:
AVG
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: O3DRIV
MODE:
AVG
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: OFSET1
MODE:
INST
PRECISION: 1
STORE NUM SAMPLES OFF
PARAMETER: SLOPE1
MODE:
INST
PRECISION:3
STORE NUM SAMPLES OFF
PARAMETER: ZSCNC1
MODE:
INST
PRECISION: 1
STORE NUM SAMPLES OFF
Figure 6-8: Default iDAS Channels Setup
These default Data Channels can be used as they are, or they can be customized
from the front panel to fit a specific application. They can also be deleted to make
room for custom user-programmed Data Channels.