Model 400E Ozone Analyzer
Instruction Manual
Operating Instructions
P/N 04316 Rev: B 81
6.6. Analog Outputs - Electronic Configuration
To access the ANALOG I/O CONFIGURATION menu and its submenus, press the
following keys:
ANALOG I / O CONFIGURATION MENU
SAMPLE
ENTER SETUP PASS : 8
18
8 1 8 ENTR EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
From this point
forward, EXIT
returns
to the
SECONDARY
SETUP MENU
EXIT returns
to the
PRIMARY
SETUP MENU
DIAG SIGNAL I / O
NEXT ENTR EXIT
DIAG ANALOG OUTPUT
PREV NEXT ENTR EXIT
DIAG ANALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
EXIT returns
to the main
SAMPLE
display
Analyzer enters analog I / O
confi
guration mode. See Table 6
-
8