EasyManua.ls Logo

Teledyne 400E - Table 3-2: Inlet Outlet Nomenclature

Teledyne 400E
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting Started Model 400E Ozone Analyzer
Instruction Manual
30 P/N 04316 Rev: B
Table 3-2: Inlet Outlet Nomenclature
Rear Panel Label Function
Sample
Connect a gas line from the source of Sample Gas here.
Calibration gasses are also inlet here on instruments without
Zero/Span/Shutoff Valve Options installed.
Exhaust
Connect an exhaust gas line of not more than 10 meters long here.
Span
On Instruments with Zero/Span Valve Options installed, connect a gas
line to the source of Calibrated Span Gas here.
Zero Air
On Instruments with Zero/Span Valve Options installed, connect a gas
line to the source of Zero Air here.
Dry Air
On instruments with Internal Zero Air Options installed attach a gas
line to the source of Dry Air here (< -20°C dew point).
2. Attach a sample inlet line to the sample inlet port. Ideally, the pressure of the
sample gas should be at ambient pressure.
The SAMPLE input line should not be more than 2 meters long.
Figure 3-3 depicts the pneumatic flow for this analyzer in its basic
configuration.
Figure 3-4 depicts the pneumatic flow for this analyzer with the
Zero/Span/Valve Option installed.
Some applications, such as EPA monitoring, require multipoint calibration checks
where Span gas of several different concentrations is needed. We recommend using
a Gas Dilution Calibrator such as a T-API Model 700 with internal photometer
option.
NOTE
Maximum pressure of gas at the Sample Inlet should not
exceed 1.5 in-Hg above ambient.

Table of Contents

Related product manuals