Model 400E Ozone Analyzer
Instruction Manual
Operating Instructions
P/N 04316 Rev: B 153
6.12.3. Editing the Data Channel Trigger Event
To edit the list of Data Parameters associated with a specific Data Channel, press:
SETUP X.X EVENT:ATIMER
<PREV NEXT> ENTR EXIT
Edit Data Channel Menu
PREV NEXT INS DEL
EDIT
PRNT
EXIT
FROM THE DATA CHANNEL PASSWORD MENU
Exits to the Main
Data Acquisition
Menu
Press each Key repeatedly to cycle through the
list of available Trigger Events (see Appendix F
for a complete list).
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X EVENT:ATIMER
<SET SET> EDIT PRINT EXIT
ENTR accepts the new string
and returns to the previous
menu.
EXIT ignores the new string
and returns to the previous
menu.