Specifications, Agency Approvals, and
Warranty
Model 400E Ozone Analyzer
Instruction Manual
18 P/N 04316 Rev: B
Table 2-2: Model 400E Internal O
3
Generator Specifications
Maximum Concentration 1.0 PPM
Minimum Concentration 0.050 PPM
Initial Accuracy +/- 5% of target concentration
Stability (7 Days) 1% of reading
Repeatability (7 days) 1% of reading
Response Time < 5 min to 95%
Resolution 0.5 ppb
Table 2-3: Specifications for Model 400E IZS Generator w/o Reference
Feedback Option
Maximum Concentration 1.0 PPM
Minimum Concentration 0.050 PPM
Initial Accuracy +/- 10% of target concentration
Stability (7 Days) 2% of reading
Repeatability (7 days) 2% of reading
Response Time < 5 min to 95%
Resolution 0.5 ppb
2.2. EPA Equivalency Designation
Advanced Pollution Instrumentation, Inc., Model 400E Ozone Analyzer is designated
as Equivalent Method Number EQOA-0992-087 as defined in 40 CFR Part 53, when
operated under the following conditions:
1. Range: Any range from 100 ppb to 1 ppm.
2. Ambient temperature range of 5 to 40°C.
3. Line voltage range of 105 – 125 and 200 – 240 VAC, 50/60 Hz.
4. With 5-micron PTFE filter element installed in the internal filter assembly.
5. Sample flow of 800 ± 80 cc/min at sea level.
6. Internal or External sample pump.

Table of Contents