Operating Instructions Model 400E Ozone Analyzer
Instruction Manual
70 P/N 04316 Rev: B
6.4.2. Setting the O
3
Generator Low-Span (Mid Point) Output
Level
For applications where To set the ozone LO SPAN (Midpoint) concentration for the
IZS O
3
generator, press:
ignores the
new setting.
ENTR accepts
new setting.
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST CAL CALZ CALS SETUP
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
NEXT JUMP EDIT PRNT EXIT
SPAN CAL M 3) O3_GEN_LOW1=100.0 PPB
0 1 0 0 .0 ENTR EXIT
Toggle these keys to change setting
•
Only Values from 0 to 1500 will be accepted..
•
A value of 0 turns the lamp OFF.
•
The
ENTR
key will disappear if an invalid setting is
SETUP X.X
CFG ACAL DAS RNGE PASS CLK
SETUP X.X
SETUP X.X ENTER VARS PASS:818
Repeat pressing
NEXT
until …
NEXT JUMP EDIT PRNT EXIT
Sets LOW SPAN Point for
RANGE1.
To Set the LOW SPAN point
for RANGE2 in DUAL or
AUTO range modes …
Press NEXT key once more
to select O3_GEN_LOW2
then continue as shown.