Operating Instructions Model 400E Ozone Analyzer
Instruction Manual
70 P/N 04316 Rev: B
6.4.2. Setting the O
3
Generator Low-Span (Mid Point) Output
Level
For applications where To set the ozone LO SPAN (Midpoint) concentration for the
IZS O
3
generator, press:
EXIT
ignores the
new setting.
ENTR accepts
the
new setting.
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST CAL CALZ CALS SETUP
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
NEXT JUMP EDIT PRNT EXIT
SPAN CAL M 3) O3_GEN_LOW1=100.0 PPB
0 1 0 0 .0 ENTR EXIT
Toggle these keys to change setting
•
Only Values from 0 to 1500 will be accepted..
•
A value of 0 turns the lamp OFF.
•
The
ENTR
key will disappear if an invalid setting is
attempted.
SETUP X.X
CFG ACAL DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
COMM
VARS
DIAG
O3 HALT
SETUP X.X ENTER VARS PASS:818
8 1 8
ENTR EXIT
Repeat pressing
NEXT
until …
SETUP X.X
3) O3_GEN_LOW1=
100.0 PPB
NEXT JUMP EDIT PRNT EXIT
Sets LOW SPAN Point for
RANGE1.
To Set the LOW SPAN point
for RANGE2 in DUAL or
AUTO range modes …
Press NEXT key once more
to select O3_GEN_LOW2
then continue as shown.