EasyManua.ls Logo

Teledyne 400E - Electrical Connections

Teledyne 400E
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 400E Ozone Analyzer
Instruction Manual
Getting Started
P/N 04316 Rev: B 25
8. VENTILATION CLEARANCE: Whether the analyzer is set up on a bench or
installed into an instrument rack, be sure to leave sufficient ventilation
clearance.
Area
Minimum Required
Clearance
Back of the instrument 4 in.
Sides of the instrument 1 in.
Above and below the instrument 1 in.
Various rack mount kits are available for this analyzer. See Section 5.1 of this
manual for more information.
3.1.1. Electrical Connections
CAUTION
Check the voltage and frequency label on the rear panel of the
instrument (See Figure 3-2) for compatibility with the local
power before plugging the M400E into line power.
CAUTION
Power connection MUST have functioning ground connection.
1. Refer to Figure 3-2 for the location of the rear panel electrical connections.
2. Attach a strip chart recorder and/or data-logger to the appropriate contacts of
the Analog Out connecter on the rear panel of the analyzer.
ANALOG OUT
A1 A2 A3 A4
+ - + - + - + -
The A1 and A2 channels output a signal that is proportional to the O
3
concentration
of the Sample Gas.

Table of Contents