EasyManua.ls Logo

Teledyne 400E - Page 41

Teledyne 400E
318 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 400E Ozone Analyzer
Instruction Manual
Getting Started
P/N 04316 Rev: B 41
3. Perform the Zero/Span Calibration Procedure:
SAMPLE RANGE = 500.0 PPB 03 =
< TST TST > CAL SETUP
M-P CAL RANGE = 500.0 PPB 03 =
< TST TST > ZERO CONC EXIT
EXIT
to Return to the Main
SAMPLE
Display
M-P CAL RANGE = 500.0 PPB 03 =
< TST TST > ENTR CONC EXIT
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the
SPAN
button
is displayed
ACTION:
Allow Zero Gas to enter
the sample port on the
rear of the instrument.
ACTION:
Switch gas streams to
span gas.
M-P CAL RANGE = 500.0 PPB 03 =
< TST TST > SPAN CONC EXIT
M-P CAL RANGE = 500.0 PPB 03 =
< TST TST >
ENTR
SPAN CONC
EXIT
M-P CAL RANGE = 500.0 PPB 03 =
< TST TST > ENTR CONC EXIT
WARNING!
Pressing the ENTR Key
changes the calibration
equations as well as
OFFSET
&
SLOPE
values for the
instrument.
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the ZERO
button
is displayed
WARNING!
Pressing the ENTR Key changes the
calibration equations as well as
OFFSET
&
SLOPE
values for the
instrument.
NOTE:
In certain insta
nces where low Span gas
concentrations are present,
both the ZERO & SPAN buttons may
appear simultaneously
If either the ZERO or
SPAN buttons fail to
appear see Section 9 for
troubleshooting tips.
4. The Model 400E Analyzer is now ready for operation.

Table of Contents