Operating Instructions Model 400E Ozone Analyzer
Instruction Manual
68 P/N 04316 Rev: B
To access the VARS menu press:
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST TST > CAL SETUP
SETUP X.X
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
COMM VARS DIAG HALT EXIT
SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT JUMP EDIT PRNT EXIT
SETUP X.X 1 ) PHOTO_LAMP=58.0 DegC
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X.X 3 ) O3_GEN_LOW1=100.0 PPB
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X ENTER VARS PASS: 818
8 1 8 ENTR EXIT
SETUP X.X 2 ) O3_GEN_LAMP=48.0 DegC
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X.X DAS_HOLD_OFF=15.0 Minutes
1 5 .0 ENTR EXIT
SETUP X.X 3 ) O3_GEN_LOW1=100.0 PPB
0 1 0 0 .0 ENTR EXIT
SETUP X.X 4 ) O3_GEN_LOW2=100.0 PPB
PREV NEXT JUMP EDIT PRNT EXIT
SETUP X.X 5 ) O2_SCRUB_SET=110.0 DegC
PREV NEXT JUMP EDIT PRNT EXIT
DO NOT
change
theses set-
points
unless
specifically
instructed to by
T-
API Customer
Service
SETUP X.X CLOCK_ADJ=0 Sec/Day
+ 0 0 ENTR EXIT
SETUP X.X 6 ) CLOCK_ADJ=0 Sec/Day
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X 3 ) O3_GEN_LOW2=100.0 PPB
0 1 0 0 .0 ENTR EXIT
Toggle these keys to change setting
EXIT
ignores the
new setting.
ENTR accepts the
new setting.
Toggle these keys to change setting
Toggle these keys to change setting
Toggle these keys to change setting