15. SPECIAL MODE (MENU KEY OPERATION)
15-7
15.2 PROGRAMS
15.2.1 TEACH: Setting functions of Teach Measurement
1) SET POINT: Set point
1-1) WRITE: Input of measuring (control) point
Input measuring point by sighting the points one by one with AP-L1A to memorize
in Teach Measurement mode. See Chapter 8.2 "Teach measurement " how to
operate.
Setting range
1point measuring points 50points
1-1-1) rl: Setting telescope inversion measurement (Set each measuring point)
Selecting items Set items
OFF
Measure at the normal position (direct) of telescope only.
ON
Measure at the normal (direct) and reverse position of telescope.
Initial value:OFF
1-1-2) MEAS: Setting of measurement times (Set each measuring point)
This is to set measuring times for each single sight.
Setting times
1 Measuring time 100
Initial value:1
1-2) DELETE: Deleting a measuring point
Delete useless measured point among measuring points. See Chapter 8.2 "Teach
measurement" how to operate.
2) PARAMETERS: Setting Parameter for Teach measurement
2-1) SEARCH: Setting SEARCH mode
When collimating a prism on memorized point and AP-L1A losses the prism (in
the case where the prism is not within the scanning area), the Search function
will be activated.
Selecting items Set items
ON Search function ON
Search area and pattern are set by special mode (menu key operation).
Measuring goes to next step when prism is not found after searching all
the search area.
OFF Search function
OFFMeasuring goes to next step when prism is not found during wait
time (WAIT) which is set by special mode (5 seconds in setting Hold).
Initial Value:ON
Note : When setting search On, be sure that there is not another
target or prism within the search area.