131
7 CHECK AND ADJUSTMENT
7
Tu rn the telescope in reverse telescope setting.
8
Collimate target A.
9
Press the [SET] key ten times.
The number of measurements is displayed at
the top right of the display.
10
Collimate target B (more than ±10
°
from the
level) in reverse telescope setting.
11
Press the [SET] key ten times.
The number of measurements is displayed at
the top right of the display. *1)
12
Tu rn the telescope in normal telescope setting.
13
Collimate target B.
14
Press the [SET] key ten times.
Then the display returns to 3AXIS
compensation menu.
*1) Pressing the [SKIP] key enables to set next step without changing the last compensated value.