Force
FXTM-C
Service
Manual
Calibrating
the
Force
FX-C
Generator
h.
Press
the
Med
(Fulgurate)
button.
Verify
that
an
"H"
appears
in
the
Cut
display
and
a "10"
appears
in
the
display.
Default
fulgurate
is
completed.
5.
To
check
the
Spray
mode,
press
the
High
(Spray)
button.
a. Verify
that
10
appears
in
the
Coag
display.
h.
Press
the
coag
pedal
anct
while
activating
the
generator,
press
the
Coag
up
(
Ll
)
or
dov,.'Il ( V )
arrow
buttons
until
the
voltmeter
reading
is
equivalent
to
141
± 13
rnA
rms.
Ignore
ch<mges to
the
Coag
display.
c.
Stop
activation.
Verify
that
the
Coag
display
changed
to 120.
d.
Press
the
coag
pedal
again
and,
while
activating
the
generator,
press
the
Coag
up
(Ll )
or
dov\'Il
(V)
arrow
buttons
until
the
voltmeter
reading
is
equivalent
to
489 6 rnA
rms.
Ignore
changes
to
the
Coag
display.
e.
Stop
activation. Verify
that
the
Coag
display
changed
to
10.
6.
To
save
the
ECON
calibration
values,
press
the
Bipolar
up
arrmv
(
Ll
)
button.
To
leave
Calibration
mode,
refer
to
Exiting Calibration
lv10d1'.
5-39
- .