Responding
to
System
Alarms
Number
81
100
101
102
103
104
105
110
111
6-12
Description
Internal
component
malfunction.
Optoisolator
test
failed.
Software
malfunction.
rem_update_state
value
outside
state
range.
Software
malfunction.
rem_pad_state
value
outside
state
range.
Software
malfunction.
rem_state
value
in
re_rem_control(
)
outside
state
range.
Software
malfunction.
rem_state
value
in
re_non_rem_control(
)
outside
state
range.
Software
malfunction.
rem_flash_state
value
outside
state
range.
Software
malfunction.
remJed_state
value
outside
state
range.
Software
malfunction.
caLstate
value
in
ca_generatocsetup(
)
outside
state
range.
Software
malfunction.
caLstate
value
in
ca_clock_date_setup(
)
outside
state
range.
Recommended
Action
Do
not
attempt
to
use
the
generator.
Record
the
number
and
call
the
Valleylab
Service
Center.
Turn
off,
then
turn
on
the
generator.
If
the
alarm
number
reappears,
record
the
number
and
call
the
Valleylab
Service
Center.
Force
F)(TM-C
Service
Manual

Table of Contents

Related product manuals