Responding
to
System
Alarms
Number
Description
Recommended
Action
164-166
Dosage
error
while
activating
a
bipolar
Repeat
the
failing
calibration
step.
If
the
number
reappears,
mode
or
a
cut
mode.
record
the
number
and
call
the
Valleylab
Service
Center.
170
Watchdog
malfunction.
Correct
value
not
Refer
to
Correctlf7g
IC
UB
Malfunctions
in
th
is
section.
sent
from
feedback
microcontroller.
171
Watchdog
malfunction.
Correct
value
not
Refer
to
Correcting
IC
U3
Malfunctions
in
this
section.
sent
from
main
microcontroller.
172
Watchdog
malfunction.
Main
microcontroller
unable
to
access
watchdog
semaphore.
173
Watchdog
malfunction.
Feedback
Refer
to
Correcting
IC
UB
Malfunctions
in
this
section.
microcontroller
unable
to
access
watchdog
semaphore.
174
Software
malfunction.
which
..
errors
value
Turn
off,
then
turn
on
the
generator.
If
the
alarm
number
a
c:
outside
state
range
reappears,
record
the
number
and
call
the
Valleylab
Service
0-
GenteL
m
.(J)
:::r
0
180
Internal
diagnostics.
Main
microcontroller
Refer
to
Correcting
IC
U3
Malfunctions
in
this
section.
0
:::t.
WSI
RAM
check
failed.
::::l
co
181
Internal
diagnostics.
Feedback
Refer
to
Correcting
IC
UB
Malfunctions
in
this
section.
microcontroller
WSI
RAM
check
failed.
182
Internal
diagnostics.
Main
microcontroller
Refer
to
Correctlf7g
IC
U3
Malfunctions
in
this
section.
dual-port
RAM
check
failed.
183
Internal
diagnostics.
Main
microcontrolier
Page
0
(zero)
ROM
CRC
test
failed.
184
Internal
diagnostics.
Main
microcontroller
Page
FROM
CRC
test
failed.
185
Internal
diagnostics.
Feedback
Refer
to
Correcting
IC
U6
Malfunctions
in
this
section.
microcontroller
ROM
GRC
test
failed.
189
Software
malfunction.
display_state
value
Turn
off,
then
turn
on
the
generator.
If
the
alarm
number
outside
state
range.
reappears,
record
the
number
and
call
the
Valleylab
Service
Genter.
Force
FXTM-C
Service
Manual
6-15