EasyManua.ls Logo

Valleylab Force FX-C - Page 136

Valleylab Force FX-C
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Responding
to
System
Alarms
Number
151
154
161
163
6-14
Description
Master
microcontrolier
unable
to
access
GEN_SEM
semaphore.
Main
microcontrolier
unable
to
access
FEEDBACK_SEM
semaphore.
Dosage
error
while
generator
was
activated,
where
Vsns
and
Isns
stay
the
same
or
decrease
while
the
EGON
increases
consistently
for
about
150
ms
..
Dosage
error
while
activating
a
coag
mode
..
Recommended
Action
Refer
to
Correcting
IC
U6
Malfunctions
in
this
section.
Refer
to
Correcting
IC
U6
Malfunctions
in
this
section.
Do
not
attempt
to
use
the
generator.
Record
the
number
and
cali
the
Valleylab
Service
Center.
Force
FXTM-C
Service
Manual

Table of Contents