EasyManua.ls Logo

Waters 2489 - Threshold Events

Waters 2489
234 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the operator interface 3-39
Tip:
If the detector is under the control of the Waters 717plus
Autosampler or another external device, the Inject Start
programmed from that device runs the method.
If you are working in real time under current conditions (method *),
at a power failure or shutdown you lose all timed or threshold
events if you do not store them as a method. See Storing a method”
on page 3-40.
Threshold events
You can program threshold events on both channel A and channel B to control
the switch contact closure outputs, for example, when using a fraction
collector. You can program the switch to change when the programmed output
(absorbance, ratio, energy, etc.) on the detector channel A or B is above a
specified threshold. Below the specified threshold, the switch is set as shown
in table below.
To define the pulse period or the frequency of a wave, see “Configuring the
detectoron page 3-22.
Threshold events “Set” parameters
Number Event
1Set switch 1
2Set switch 2
Threshold events “To” parameters
Number Set To
Below Threshold
Switch State
1On Off
2Off On
3Pulse Off
4Rect wave
(rectangular
wave)
Off

Table of Contents

Other manuals for Waters 2489

Related product manuals