Scanning spectra 3-45
The figure below shows two scans of anthracene overlaid. At a pace of
1000 nm/min, the overlaid scan (with dotted line) shows a reduced
number of points scanned, lowering the resolution in comparison with
the original scan done at a pace of 100 nm/min.
Scan of anthracene at 100 nm/min and 1000 nm/min
Tip: The higher the number you enter in the Pace field, the lower the
resolution of the scan.
Pace and sampling resolution examples
Pace (in nm/min) Sampling Resolution (in nm)
100 and less 0.5
200 1.0
400 2.0
Anthracene scan at 100 nm/min and 1000 nm/min
Absorbance
Wavelength