3-46 Preparing the Detector
Tickmarks – This value allows tickmarks to be generated at the
specified wavelength increment to assist in the interpretation of charted
data.
The figures below show a 190 to 600-nm scan of erbium standard in a
cuvette at a pace of 200 nm/min with tickmarks specified every 20 nm
and the same scan without tickmarks.
AUFS – The sensitivity setting for scaling the charted spectrum.
Scan of erbium standard in a cuvette from 190 nm to 600 nm at a pace
of 200 nm/min with tickmarks every 20 nm
Scan of erbium standard in a cuvette from 190 nm to 600 nm at a pace
of 200 nm/min without tickmarks
Enter these parameters when you select the type of scan, either zero or
sample.
Wavelength (in nm)
290 390 490
AU
0.0
0.1
0.2
0.3
0.4
0.5
0.6
Wavelength (in nm)
AU
0.0
0.1
0.2
0.3
0.4
0.5
0.6
290 390 490