EasyManua.ls Logo

XITRON XT2640 - Voltage Measurement Accuracy

Default Icon
187 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XT2640OperatingManual 13July2016 Page167of187
28.2.2 VOLTAGE MEASUREMENT ACCURACY
ThechartsbelowshowguaranteedmaximumvoltageerrorsforDC,MAINS,AVIONICS,and50kHzthroughouta1Vto1000Vrange
ofappliedvoltagesexpressedas%ofreadingandarevalidwithin±5Cofthecalibrationtemperature(add0.005%perCbeyond
this)andwherenosignificantcommon‐modeispresent.Followingthechartsisatablewhichcanbeusedtocalculatethe
guaranteedaccuraciesforapplicationsotherthanshowninthechartsandalsoforthecomputationofnumericalerrors.
0.01%
0.10%
1.00%
1 10 100 1,000
DCVoltageAccuracy
LChannel
SChannel
AChannel
WChannel
0.01%
0.10%
1.00%
1 10 100 1,000
MAINSVoltageAccuracy
LChannel
SChannel
AChannel
WChannel
0.01%
0.10%
1.00%
1 10 100 1,000
AVIONICSVoltageAccuracy
LChannel
SChannel
AChannel
WChannel
0.1%
1.0%
10.0%
1 10 100 1,000
50kHzVoltageAccuracy
LChannel
SChannel
AChannel
WChannel

Table of Contents