XT2640OperatingManual 13July2016 Page24of187
9.5 2Ø3WOR3Ø3W(2CH)WIRINGUSINGEXTERNALCURRENTMEASUREMENT
CAUTION:
FORANA:ATYPECT:USEWITHSH,SD,AH,AD,LH,LD,WHORWDCHANNELSONLY
FORANA:VTYPECTORASHUNT:USEWITHSX,AX,LXORWXCHANNELSONLY
DONOTCONNECTWIRINGWHILEVOLTAGEPRESENT.
ENSURE
WIRINGISCORRECTBEFOREAPPLYINGVOLTAGE.
WIRING:
Thickerconnectionsarecurrentcarrying.Ensurewireusedisratedforthemaximumvoltageandcurrenttobe
encountered.
Thinnerconnectionsarenotcurrentcarrying.Ensurewireusedisratedforthemaximumvoltagetobeencountered.
Phasesareshownfordescriptivepurposes,theXT2640placesnorelianceontheactualphasinghoweverusingthose
shownensuresthattheXT2640displaymatchestheactualphasingifthechannelshaveincreasingchannelnumbersfrom
lefttorightinthediagramabove.
ManyCTdevicesareisolating,inthosecasesyoushouldgroundtheA‐(BLUE)terminalofthechanneliftheCTdoesnot
groundit.
XT2640CONFIGURATION:
IntheXT2640MEASCONFIGmenuyoushouldappropriatelysetSCALEforeachchanneltosuittheCTorshuntbeingused
(bothascalefactorandanoffsetareavailable).
A
V
XT2640
C
H
CT
+ -
øA
From
SOURCE
N or øC
From
SOURCE
øA
To
LOAD
A
V
XT2640
CH
CT
+ -
øB
From
SOURCE
øB
To
LOAD
N or øC
To
LOAD