XT2640OperatingManual 13July2016 Page81of187
TheFUNDsettingoftheDCVPAshouldbesettotheVPAmeasuringthePWMGenerator.
TheHARMONICSsettingfortheDCVPAshouldbesetto0(itcanalsobesettoanon‐zerovalue,inwhichcasethe
ripplecontentoftheDCsupplywillbeavailableasharmonics).
UsethesameLF/PERIODandRESPONSEsettingsinbothVPAs.
InthismannerthetwoVPAmeasurementswillbecloselysynchronizedbyalwaysstartingthematthesametimeso
efficiencywilltrackfastchangestosignals.Bothmeasurementperiodswillbealmostthesame,sosynchronizationwillbe
verycloseandmeasurementswillbealmostcontinuousasanypausesbetweenmeasurementswillbeveryshort(typically
afewtensofµsec).
3.
Iftherewillbesignificantchangestothelevelsorsignalswhichrequiretrackingforefficiencyresultsandthesupplyis
eitherDCcontainingrippleatsomefrequencyotherthanthemodulationfrequency,oritisanACsupply(e.g.mains).
TheMEASMODEshouldbesettoSYNCVPA.
BothsetsofmeasurementsshouldbeconfiguredinseparateVPAs.AnyVPAcanbeusedforeithermeasurementand
bothareindependentlyconfigured.Itisrecommendedthat,ifpossible,bothLF/PERIODsettingsbethesamevalue.
UsethesameRESPONSEsettinginbothVPAs.
InthismannerthetwoVPAmeasurementswillbecloselysynchronizedbyalwaysstartingthematthesametime,even
thoughtheymightnotactuallyendatexactlythesametime,soefficiencywilltrackfastchangestosignalsasmuchas
possiblebuttheremaybeaslightdifferencebetweenthetwomeasurementperiodsandneithermeasurementwillbe
continuousaseachmustwaitfortheothertofinishbeforeitcanrestart.
4.
Ifnoneoftheabovesituationsaretrue,orifyouareuncertainifanyoftheabovesituationsaretrue.
TheMEASMODEshouldbesettoMULTIVPA.
BothsetsofmeasurementsshouldbeconfiguredinseparateVPAs.AnyVPAcanbeusedforeithermeasurementand
botharecompletelyindependentlyconfigured.
Althoughthetwomeasurementperiodswillbeasynchronousyoucanreduceanyabnormalefficiencyresultsduring
fastchangesasfollows–
o UsethesameLF/PERIODsettinginbothVPAsifpossible.
o UsetheshortestpossiblemeasurementperiodintheLF/PERIODsettingsofbothVPAs.
o UsethesameRESPONSEsettinginbothVPAs.
InthismannerthetwoVPAmeasurementswillbenotbesynchronizedinanymanner,soefficiencymaynotcorrectlytrack
fastchangestosignalsbutwillbeaccuratewhilebothmeasurementsarestable.Inthisoptionbothmeasurementsare
alwayscontinuous.Notethatyoucanreadilychangetooption#3abovebysimplychangingtheMEASMODEtoSYNCVPA.