6-8 IM 765501-01E
Keep
The last source level of the sweep operation is retained even after the sweep operation
is complete.
Sd
Trigger
Meas
Trigger Trigger Trigger Trigger
Sd
Meas Sd Meas Sd Meas Sd Meas
Source measure
cycle
Source level
Stop value
Start value
Source level
before sweep
Sd: Source delay
Meas: Measurement delay + measurement + computation
Waiting for
start
Sweep state
Start
Waiting for
start
Sweeping
• Sweep operation when the termination mode (Last) is set to keep
<<Corresponding Command Mnemonic>>
:SWEep:LAST
6.3 Selecting the Termination Mode