EasyManua.ls Logo

YOKOGAWA 765501 - Page 110

YOKOGAWA 765501
277 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-5
IM 765501-01E
Measurement
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
18
17
16
Index
App
Explanation
Measurement Delay
You can set the wait time at which the measurement is started after the source level is
switched.
Trigger
Int or Ext
Source
Measurement
Source measure cycle
Measurement
time
Measurement delay
Source delay
Source value
Computation
Selectable Range
1 µs to 3600.000000 s
Note
The time needed for the source value to stabilize after the source is applied varies depending
on the load, source value, and limiter value. Adjust the measurement delay by taking these
into account as well as the time required for the measurement accuracy test.
<<Corresponding Command Mnemonic>>
:SENSe:DELay
7.4 Setting the Measurement Delay

Table of Contents

Related product manuals