16-56
IM 765501-01E
‘—————————————————————————————————
‘ Generate triggers and read the measured results
‘ Generate triggers from a PC and read the results
‘—————————————————————————————————
Function TrigReadAsString(dev As Integer) As String ‘ String output
TrigReadAsString = CommQuery(dev, “:INIT;*TRG;:FETC?”)
End Function
Function TrigReadAsDouble(dev As Integer) As Double ‘ Real number output
TrigReadAsDouble = Val(TrigReadAsString(dev))
End Function
‘—————————————————————————————————
‘ Sample Sequence 2
‘ Set the GS610 to Setup 2 and vary the source level as follows: 2.8 V, 2.9 V,
‘ 3.0 V, 3.1 V, and 3.2 V. Then, generate a trigger, read the results, and turn
‘ the output OFF.
‘—————————————————————————————————
Function SampleSequence2()
Dim dev As Integer
Dim result(5) As Double
dev = CommOpen() ‘ Open the device
Call Setup2(dev) ‘ Set to Setup 2
Call CommSend(dev, “:SOUR:VOLT:LEV 2.8”) ‘ Set the level to 2.8 V
result(0) = TrigReadAsDouble(dev) ‘ Generate a trigger and read the
‘ result
Call CommSend(dev, “:SOUR:VOLT:LEV 2.9”) ‘ Set the level to 2.9 V
result(1) = TrigReadAsDouble(dev) ‘ Generate a trigger and read the
‘ result
Call CommSend(dev, “:SOUR:VOLT:LEV 3.0”) ‘ Set the level to 3.0 V
result(2) = TrigReadAsDouble(dev) ‘ Generate a trigger and read the
‘ result
Call CommSend(dev, “:SOUR:VOLT:LEV 3.1”) ‘ Set the level to 3.1 V
result(3) = TrigReadAsDouble(dev) ‘ Generate a trigger and read the
‘ result
Call CommSend(dev, “:SOUR:VOLT:LEV 3.2”) ‘ Set the level to 3.2 V
result(4) = TrigReadAsDouble(dev) ‘ Generate a trigger and read the
‘ result
Call CommSend(dev, “:OUTP:STAT OFF”) ‘ Turn the output OFF
Call CommClose(dev) ‘ Close the device
End Function
16.4 Sample Programs

Table of Contents

Related product manuals