Explanation
Zero Generation
When generating voltage, 0 V is generated. When generating current, 0 A is generated.
Setting the Zero Offset
An offset can be specified on the zero value when generating zero output.
Selectable Range of Zero Offset
–110.000 V to 110.000 V
Note
If the zero offset value exceeds the source range of the selected range setting, the zero offset
is set to the upper or lower limit of the range setting.
Setting the Zero Source Impedance
The impedance when generating a zero level signal can be selected.
When Low Impedance (LoZ) Is Selected When Generating Voltage or When High
Impedance (HiZ) Is Selected When Generating Current
The source value is set to 0.
When High Impedance (HiZ) Is Selected When Generating Voltage or When Low
Impedance (LoZ) Is Selected When Generating Current
The source value is set to 0, and the limiter is set to 0.5% of the range setting.
<<Corresponding Command Mnemonic>>
:OUTPut[:STATe]
:SOURce:VOLTage:ZERO:OFFSet
:SOURce:VOLTage:ZERO:IMPedance
:SOURce:CURRent:ZERO:OFFSet
:SOURce:CURRent:ZERO:IMPedance
5.8 Zero Generation