6-30
IM 701310-01E
Example
L
H
H
CH1
CH2
Trigger
CH3
CH1
CH2
H
H
L
L
L
Qualification
not met
Qualification
not met
Qualification
met
t1
t2
t1>T t2>T
t3
t3>T
L
Qualification
met
Trigger
(Qualification met)
Trigger source: CH3, More than
Qualification: CH1 = H, CH2 = L, other channels = X, AND logic
6.11 Activating a Trigger on a Conditional Pulse Width