7-12
IM 701310-01E
Setting the Triggered Actions
4.
Press the Action soft key.
5.
Press an individual action’s soft key to turn it ON or OFF.
If you turn Mail-Mode ON, set the Interval with the rotary knob.
6.
Press ESC to return to the previous screen.
Setting the Signal Acquisition Count
7.
Use the rotary knob to set the signal acquisition count.
Activating the Action-On-Trigger Function
8.
Press the EXEC soft key to begin acquiring the signal and to activate the action-
on-trigger function.
The Exec soft key changes to the Abort soft key.
Stopping the Action-On-Trigger Function
9.
Press the Abort soft key or START/STOP to stop the action-on-trigger function
and the acquisition of the signal.
7.8 Setting the Action-On-Trigger Function