7-2
IM 701310-01E
Explanation
Acquisition Modes
You can select the acquisition mode from one of the following. The default setting is
Normal.
• Normal Mode
The DL9000 saves sampled data to the acquisition memory without performing any
special data processing.
• Envelope Mode
The maximum and minimum values are determined every acquisition interval from the
data sampled at 2.5 GS/s (DL9040/DL9040L/DL9140/DL9140L) or 5 GS/s (DL9240/
DL9240L). The maximum and minimum values are stored in the acquisition memory
and an envelope waveform is displayed.
•
A
veraging Mode
The DL9000 averages the sampled data and saves it to the acquisition memory.
The
averaging method varies depending on the trigger mode.
• When the trigger mode is
Auto, Auto Level, or Normal, expone
ntial averaging is
used. Set the attenuation constant to a value from 2 to 1024.
• When the trigger mode is Single or N Single, simple averaging
is used. Set the
sampled data acquisition count to a value from 2 to 65536.
• In Single mode, set the acquisition count from the acquisition m
ode menu.
• In N Single mode, set the acquisition count from the trigger mode menu.
An = {(N – 1)An–1 + Xn}
1
N
Exponential average
(When the trigger mode is Auto, Auto Level, or Normal)
AN =
N
Simple average
(When the trigger mode is Single or N Single)
N
n = 1
Σ Xn
An:
Xn:
N:
n
th
average
n
th
measured value
Attenuation constant
(2 to 1024, in powers-of-2 increments)
Xn:
N:
n
th
measured value
Acquisition count
(2 to 65536, in powers-of-2 increments)
Notes about Averaging
• Only repeating signals are averaged.
• If the trigger is not consistent, and the waveform is not completely synchronized, the
signal will not be averaged correctly, and the waveform will be distorted.
•
Roll mode is not valid when averaging is used.
• If you stop acquisition by pressing ST
ART/ST
OP, averaging will also stop. When you
restart acquisition, values will be averaged from the beginning.
• With simple averaging, acquisition stops when the signal has b
een acquired for the
number of times set by the acquisition count.
• W
h
en a signal is acquired in averaging mode, the following data is saved to the history
memory.
•
Exponential averaging when the trigger mode is
Auto, Auto Level, or Normal
Each waveform averaged over a set time interval is saved to a separate record.
• Simple averaging when the trigger mode is Single
The DL9000 averages the set number of waveforms, and save
s the averaged
waveform to the most recent history memory record.
• Simple averaging when the trigger mode is N Single
The DL9000 saves the set number of waveforms to the history
memory and saves
the waveform that is the average of those waveforms to a separate section in the
history memory.
• The maximum length of records for which averaging is possibl
e is 625 KW for
DL9040/DL9140/DL9240, and 1.25 MW for DL9040L/DL9140L/DL9240L.
7.1 Setting the Acquisition Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040L and is the answer not in the manual?

YOKOGAWA DL9040L Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040L
CategoryTest Equipment
LanguageEnglish

Related product manuals