Trouble Shooting
124
Operation Manual
(For equipment with 3D function only)
Symptom Possible cause Action
Reconstruction of
3D image is not
completed.
3D image was acquired while a
panoramic image was open.
Acquire a 3D image after closing all
images.
3D images on multiple parts of the same
patient were consecutively acquired.
When acquiring 3D images on multiple
parts, proceed with the next image
acquisition only after checking that the
previously acquired image has been
reconstructed.
The reconstructed
3D image acquired is
distorted.
The 3D sensor is dislocated. Contact the dealer.
The 3D image acquired
is blurred.
The patient may have moved while the
image acquisition was in progress.
Fasten the patient's head using the head
bandontheheadxatorfor3Dimage
acquisition, and instruct the patient not
to move while image acquisition is in
progress.
The 3D image acquired
is out of the normal
domain.
Excessive load may have been applied to
theheadxatorfor3Dimageacquisition,
and the bite plate may be dislocated.
Instruct the patient not to lean on the
headxatorfor3Dimageacquisition.