Operation Manual
Electromagnetic Compatibility
TABLE:Guidance and manufacturer's declaration - electromagnetic emissions
TheX-eraSmartisintendedforuseintheelectromagneticenvironmentspeciedbelow.
The customer or the user of the X-era Smart should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The X-era Smart uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The X-era Smart is suitable for use in all establishments other than
domestic, and may be used in domestic establishments and those
directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes, provided the following
warning is heeded:
Warning: This equipment/system is intended for use by healthcare
professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It may
be necessary to take mitigation measures, such as re-orienting or
relocating the X-era Smart or shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltageuctuations/
ickeremissions
IEC 61000-3-3
Complies
TABLE: Guidance and manufacturer's declaration - electromagnetic immunity
TheX-eraSmartisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerorthe
user of the X-era Smart should assure that is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment –
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floors should be wood, concrete,
orceramictile.Ifoorsare
covered with synthetic material the
relative humidity should be at least
30%.
Electrical fast
transient / burst
IEC 61000-4-4
±2kVforpowersupplylines
±1kVforinput/outputlines
±2kV
±1kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kVdifferentialmode
±2kVcommonmode
±1kV
±2kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltagedips,short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0,5 cycles
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
Functions Interruption
Functions Interruption
Functions Interruption
Functions Interruption
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
X-era Smart requires continued
operation during power main
interruptions, it is recommended
that the X-era Smart be powered
from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magneticeld
IEC 61000-4-8
3 A/m Complies
Powerfrequencymagneticelds
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.