14
Operation Manual
TABLE: Guidance and manufacturer's declaration - electromagnetic immunity
TheX-eraSmartisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerorthe
user of the X-era Smart should assure that is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment –
guidance
Conducted RF
IEC 61000-4-6
3Vrms
150 kHz to 80 MHz
3V
Portable and mobile RF communications
equipment should be used no closer to
any part of the X-era Smart, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance
d=1.2√P
d=1.2√P80MHzto800MHz
d=2.3√P800MHzto2.5GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
FieldsstrengthsfromxedRF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
Radiated RF
IEC 61000-4-3
3V/m
80 MHz to 2.5 GHz
3V/m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorptionandreectionfromstructures,objectsandpeople.
a:Fieldstrengthsfromxedtransmitters,suchasbasestationsforradio(cellular/cordless),telephones
andlandmobileradios,amateurradio,AMandFMradiobroadcast,andTVbroadcastcannotbepredicted
theoreticallywithaccuracy.ToassesstheelectromagneticenvironmentduetoxedRFtransmitters,an
electromagneticsitesurveyshouldbeconsidered.Ifthemeasuredeldstrengthinthelocationinwhichthe
X-era Smart is used exceeds the applicable RF compliance level above, the X-era Smart should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the X-era Smart
b:Overthefrequencyrange150kHzto80MHz,eldstrengthsshouldbelessthan3V/m.