108  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Profibus module: Electrical connections 
 
RS485 interface 
Pin Signal  Explanation 
1 –  not assigned 
2  M24  24 V output voltage, ground 
3  RxD/TxD-P  Receive/send data plus, B line 
4 –  not assigned 
5  DGND  Data transmission potential (reference potential for VP) 
6  VP  Supply voltage plus (5 V) 
7  P24  24 V output voltage plus, max. 0.2 A 
8  RxD/TxD-N  Receive/send data N, A line 
9 –  not assigned 
Design: 9-pole female Sub-D plug 
MBP interface (not intrinsically safe) 
Pin Signal 
1 + 
2 Shield 
3 – 
4 not assigned 
Design: 4-pole plug-in terminal strip. Please refer to the information about 
the requisite material (see page 116)! 
NOTE 
For additional information on "Profibus", refer to the Technical Information 
"AO2000 Profibus DP/PA Interface".