AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  135
 
 
 
Warm-up phase 
Warm-up phase 
The duration of the warm-up phase depends on which analyzer module is 
installed in the gas analyzer. 
Analyzer module  Duration of the warm-up phase 
Caldos25 1.5 hours 
Caldos27  30/60 minutes for class 1/2 measurement ranges
 1)
 
Fidas24  ≤ 2 hours 
Fidas24 NMHC  ≤ 2 hours 
Limas11 IR  Approx. 2.5 hours 
Limas21 UV  Approx. 2.5 hours 
Limas21 HW  Approx. 4 hours 
Magnos206  ≤ 1 hour 
Magnos28  ≤ 1 hour 
Magnos27  2 to 4 hours 
Uras26  Approx. 0.5/2 hours without/with thermostat 
ZO23  Approx. 15 minutes 
1) See "AO2000 Series" data sheet for class details. 
NOTES 
The warm-up phase can take longer if the gas analyzer was not brought to 
room.  
During the warm-up phase measurement values can be outside the ranges 
specified in the data sheet. 
Warm-up phase duration 
The warm-up phase is deemed to be over when the measured value drift 
indication is acceptable. This depends on the size of the measurement 
range. 
NOTE for Fidas24 
Unused sample gas lines and sampling probes can release hydrocarbons 
after the first start-up over a longer period. Therefore, it can take a longer 
time until the measured value drift has reached an acceptable value.