AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  163
 
 
 
Filter parameterization 
Menu path 
MENU → Configure → Component-Specific → Filter → 
Select Component → ... 
Range 
0 to 60 seconds 
Procedure 
 
Parameter Explanation  Action 
Linear filter (Caldos25, Caldos27, Magnos27, Limas11 IR, Limas21 UV,   
Limas21 HW): 
T90 
Low-pass time constant  Set 
Non-linear filter (Fidas24, Magnos206, Magnos28, Uras26): 
T90-1 
Low-pass time constant for   
constant measured value 
Set 
T90-2 
Low-pass time constant for   
measured value changes 
Set 
Threshold 
Threshold. T90-2 applies for overshoot  Set 
Non-linear filter 
T90-2 should be adjusted T90-2 ≤ T90-1 for the non-linear filter.   
The switching threshold (in %) is generally based on the largest measure-
ment range selected (reference measurement range). 
Recommendation for 
Fidas24: T90-1 = 20 s, T90-2 = 1 s, threshold = 0.001 % 
Magnos206: T90-1 = 3 s, T90-2 = 0 s, threshold = 0.1 % 
Magnos28: T90-1 = 3 s, T90-2 = 0 s, threshold = 0.1 % 
Uras26: T90-1 = 5 s, T90-2 = 0 s, threshold = 0.6 %