AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  259
 
 
 
Magnos27: Notes for calibrating 
Sample components 
The Magnos27 analyzer module has at least one sample component with 
one measurement range and one measurement range for flue gas meas-
urements of O
2
 in N
2
 substitute gas component with 1 range. 
Associated gas effect 
Because of the thermomagnetic measurement technique employed by the 
Magnos27 analyzer module, associated gases have an effect on results. 
For this reason the sample gas composition must be considered during 
initial calibration at the factory. 
Test gases 
Zero calibration: Oxygen-free process gas or substitute gas 
Span calibration: Process gas with a known oxygen concentration or sub-
stitute gas, e.g. dried air 
Test gas for corrected sample components 
During calibration possible electronic cross-sensitivity and/or carrier gas 
corrections by other measurement components are switched off. Therefore, 
corrected measurement components should be calibrated only using a test 
gas consisting of the measurement component and an inert gas like Nitro-
gen. 
Calibration with substitute gas 
If test gases are not available for calibration, the analyzer module can be 
factory-set for calibration with a substitute gas (see also section "Calibra-
tion methods" (see page 223)). This setting is documented in the analyzer 
data sheet. 
Substitute gas calibration of the Magnos27 analyzer module is described 
using the example of "Oxygen measurement in flue gas" (see page 260). 
Sequence of calibration 
Zero-point calibration must always precede end-point calibration.   
Wait until the warm-up phase has ended 
The analyzer module should only be calibrated after the warm-up phase (see 
page 135).