AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  37
 
 
 
Installation location 
Installation location requirements 
The gas analyzer is only intended for installation indoors. 
The technical data of the gas analyzer (see data sheet and section "Analyzer 
module operating specifications" (see page 346)) are valid for an installation 
location altitude up to 2000 meters above sea level. 
The installation site must be stable enough to bear the weight of the gas 
analyzer! 
Short gas paths 
Install the gas analyzer as close as possible to the sampling location. 
Install the gas conditioning and calibration modules as close as possible to 
the gas analyzer. 
Adequate air circulation 
Provide for adequate natural air circulation around the gas analyzer. Avoid 
heat build-up. 
When installing several system housings in a 19-inch housing, maintain a 
minimum spacing of 1 height unit between housings. 
The complete surface of the system housing is used to dissipate the heat 
losses. 
Protection from adverse conditions 
Protect the gas analyzer from 
•  Cold, 
•  Exposure to heat from e.g. the sun, furnaces, boilers 
•  Temperature variations 
•  Strong air currents 
•  Accumulation of dust and ingress of dust 
•  Corrosive atmosphere 
•  Vibration.