AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B  353
 
 
 
Magnos27: Operating specifications 
Linearity deviation  ≤ 2 % of span 
Repeatability  ≤ 1 % of span 
Zero drift  ≤ 1 % of span per week 
Sensitivity drift  ≤ 2 % of the measured value per week 
Output fluctuation (2 
σ) 
≤ 0.5 % of smallest measurement at electronic T90 time = 0 s 
Detection limit (4 
σ) 
≤ 1 % of smallest measurement at electronic T90 time = 0 s 
Flow influence  ≤ 1 % of span at a flow change of ±10 l/h. At the same sample and test gas 
flow rate, the flow influence is automatically compensated. 
Associated gas effect  Magnos27 calibration applies only to the sample gas (= sample component + 
associated gas) shown on the identification plate. 
Temperature influence  At zero point: ≤ 2 % of span per 10 °C;   
On sensitivity: ≤ 0.5 % of measured value per 10 °C;   
each with respect to the temperature during calibration.   
Thermostat temperature = 63 °C; Ambient temperature in permissible range 
Air pressure effect  At zero point: < 0.05 vol.% of O
2
 per 1 % air pressure change;   
On sensitivity without pressure correction: ≤ 1.5 % of the measured value per 
1 % air pressure change;   
On sensitivity with pressure correction (option): ≤ 0.25 % of the measured 
value per 1 % air pressure change.   
Option: Operating altitude over 2000 m 
Power supply influence  24 V DC ± 5 %: ≤ 0,2 % of span 
Position influence  Approx. 3 % of smallest measurement range span per 1° deviation from hor-
izontal orientation. Position has no effect on the hard-mounted unit. 
T
90
 time  T
90
 = 10 to 22 s, depending on sample gas flow and sample chamber connec-
tion (applies to a gas analyzer with one analyzer module)