334  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Caldos25, Caldos27, Magnos206, Magnos28, Magnos27: 
Troubleshooting 
Temperature problem 
Connecting leads of the temperature sensor or the heater discon-
nected 
•  Check the connecting leads and connectors. 
•  Check the line seating in the insulated jackets. 
Defective thermal link 
•  Check thermal link for continuity and replace if necessary. 
Leaking thermostat or open purge gas connections 
•  Check the seal integrity between the thermostat chamber and the 
mounting flange; tighten mounting bolts or replace O-rings as needed. 
•  Seal the analyzer purge gas inlet and outlet with sealing connectors. 
•  Check the purge gas and reduce as needed: in operation max. 20 l/h, 
positive pressure p
e
 = 2 to 4 hPa. 
Unstable readings 
Gas path leakage 
•  Check the seal integrity (see page 268) of the analyzer module gas paths 
and (if applicable) of the lines to the pneumatics module.