114  AO2000 CONTINUOUS GAS ANALYZERS  |  OI/AO2000-EN REV. B 
 
 
 
Standard terminal connections 
Basic principles 
The terminal connections are allocated 
•  in the order of the registered analyzer modules and 
•  within an analyzer module, in the order of the sample components. 
The order of the analyzer modules and sample components is documented 
in the analyzer data sheet and on the type plate. Beginning with analyzer 
module 1 and sample component 1, the input and output functions are first 
of all allocated in turn to available free connections of the I/O modules (slots 
–X20 to –X29). 
Profibus, Modbus 
The slot of the optional Profibus module (see page 108) is always –X20.  
The slot of the optional Modbus module (see page 109) is –X20, or –X22 if a 
Profibus module is present. 
Analog outputs 
Analog outputs are available at the 2-way analog output module or the 
4-way analog output module (see page 110). Analog output is allocated for 
each sample component in the order of the sample components. 
Alarm values 
Alarm values are available at the Digital I/O module (see page 112) with the 
standard function block application "Status signals/ext. calibration" (if the 
gas analyzer has been set to collective status during the installation of an 
analyzer module) or on the Digital I/O module with the standard function 
block application "Alarm values". An alarm value is allocated for each sample 
component in the order of the sample components.